5.4.9 Maintenance QC
The Maintenance QC Report provides a report on drawer temperatures and blocked stations, and
provides spaces to log user verification and maintenance activities (such as verifying station and
system LED indicators). The report typically prints on one page, but if needed the information may
continue on a second page. The report provides the following information:
Header: Hospital Information (if configured), Instrument Number, Instrument Software Version,
Report Title, and Date and Time Printed
Body of Report: Instrument Serial Number, Drawer Setpoint Temperature, Drawer QC
Thermometer Reading (user entry), Internal Green/Red LED Pass/Fail (user entry), Blocked (and
unusable) Stations, External Yellow/Red/Green Drawer Indicator Pass/Fail (user entry), Audible
Alert Pass/Fail (user entry), Filters Change/Date (user entry), Comments (user entry),
Technologist/Date (user entry)
Applicable Selection Criteria:
N/A
See Figure 5-21 for a sample Maintenance QC Report.
To print a QC Report:
1 From the Status display, select the Reports tab to access the Reports menu.
2 Select the desired report in the menu.
3 Select Print to print the report.