5 – Reference
133
5.4.14 Unloaded Negative Vials
The Unloaded Negative Vials Report lists all the sequenced negative vials (out-of-protocol negative
and manual negative) that have been removed from the instrument in a specified time period and have
not been reentered. The report prints in landscape mode. The report provides the following
information:
Header: Selection/Sorting/Reporting Criteria, Hospital Information (if configured), Instrument
Number, Instrument Software Version, Report Title, and Date and Time Printed
Body of Report (columns from left to right): Accession, Patient Name (30 characters), Patient ID,
Sequence, Test*/Protocol, Status*, State (column not labeled, body of report shows: ? for
anonymous, vial icon for current, or blank for removed/pending), Service, Location, TIP/TTD (Time
in Protocol for negatives and manual positives, Time To Detection for positives), “was positive”
indicator (plus sign [+] at end of line indicates that a vial that is not currently Positive previously had
a positive result)
* see Section 11 for reported values
Applicable Selection Criteria:
Time Range Sort By Report By
See Figure 5-26 for a sample Unloaded Negative Vials Report.
To print an Unloaded Negative Vials Report:
1 From the Status display, select the Reports tab to access the Reports menu.
2 Select the desired report in the menu.
3 Select the desired report criteria by selecting the radio button next to the criteria.
Time Range: Select the From: and To: buttons to select the start and end times. The default
time range is from midnight yesterday.
Sort By: Choose 1st and 2nd Sort By field of either: Accession, Hospital Service, Media
(alphabetic), Patient ID, Patient Name, Sequence, Start of Protocol, Location, Status, or None.
The default sort order is by Accession, then by Sequence.
Report By: primary selection: None, Media, Hospital Service. If a criteria other than None is
selected, any Media or Hospital Services that have been used in patient/specimen records
appear in the secondary selection field. The default values are All Media and Hospital
Services.