BD BACTEC™ FX Instrument User’s Manual
88
Positive Removal Buttons:
Verified button
Appears only if the vial sequence number was entered manually. Select to confirm to the
system that the displayed vial sequence number is correct.
Wrong button
Appears only if the vial sequence number was entered manually. Select to inform the system
that the displayed vial sequence number is incorrect. Message WE06 then appears (see
Section 7). Related vials are not shown if you select the wrong button.
Select to exit the display and return to the Status display.
5.3.2 Negative Removal Display
The Negative Removal display appears when you initiate the remove negative vials activity, by
scanning a negative vial barcode sequence number, removing a negative or manual negative vial from
a station, or by selecting Remove Negatives. It provides information about the negative vial that was
scanned.
You are informed of negative vials by the following:
GREEN System indicator illuminated
Station indicators: FLASHING GREEN
Remove Negative vials indicator/button on Status display
Negative counter on activity displays
See Figure 5-3 for a sample negative removal display.
To access Negative Removal:
Open a drawer and remove a negative vial, or
Negative Removal Fields:
Accession
Read-only field that shows the accession number of the vial.
Sequence
Read-only field that shows the vial barcode sequence number.
Medium
Read-only field that shows the medium type.
TIP
Read-only field that shows the Time in Protocol, in days; hours; minutes. Time in Protocol is
calculated as the amount of time between the vial’s Start of Protocol and the current time.
Exit button
• Open a drawer and select Remove Negatives on the Status display