BD BACTEC™ FX Instrument User’s Manual
174
Workflow Exception Messages
WE02
Unexpected vial
pulled. Remove?
You have pulled a vial that
doesn't correspond to the
current activity (e.g., pulling a
vial that is not positive during
Positive Removal, pulling any
vial during Vial Entry, etc.).
The sequence, station, and
status are shown at the top of
the message box.
Select Yes to remove the scanned vial
named in the message box.
Select No to place the vial back in the
instrument. WE03: Scan sequence to
return or select Cancel to accept
removal message then appears.
WE03
Scan sequence to
return or touch
Cancel to accept
removal.
Appears if you respond No to
WE02: Unexpected vial
pulled. Remove?
To return the vial to the instrument,
scan the vial sequence number and
place the vial in an available station.
To remove the vial, select Cancel.
WE04
Unexpected
sequence scanned.
Can you scan
correct sequence?
Appears if you scan an
unexpected vial after a
WE03: Scan sequence to
return or select Cancel to
accept removal (e.g., you
remove a positive vial then
inadvertently scan another
vial’s sequence number).
Appears if you scan an
unexpected sequence during
an activity; the sequence
scanned does not match the
sequence in the database for
the station/vial.
If you select Yes, then the WE03: Scan
sequence to return or select Cancel to
accept removal message reappears.
If you select Cancel, then a WE06:
Unverified sequence. Return through
Vial Entry workflow message appears.
To return the unexpected vial, select
the Yes button in this message.
WE05
Sequence was
manually entered.
Visually verify for
return. Verified?
The vial sequence number of
the vial being removed and/or
entered was entered
manually via the onscreen
keyboard.
The sequence is shown at the
top of the message box.
Compare the actual vial sequence
number to the one shown at the top of
the message box. If the two numbers
are identical, select Yes. If the
numbers are not identical, select No.
WE06
Unverified
sequence. Return
through Vial Entry
workflow. Consult
Manual.
Appears if you respond No to
WE04 or Cancel to VR02.
Also occurs if you respond
wrong when verifying a
manually entered sequence
number.
When the current activity is complete,
use the Vial Entry activity to enter the
vial into the instrument. Note any
additional messages that appear at
that time about vial status.
WE07
The instrument has
lost connectivity to
the server
database. Vial may
only be returned
anonymously.
An ongoing vial was pulled
from an instrument in
degraded mode in a
BD
EpiCenter configuration.
Return the vial to the station from
which it was removed to continue to
test the vial anonymously.
Identify the vial when communications
with BD EpiCenter are reestablished.
Error
No.
Message Possible Cause(s) Corrective Action(s)