7 – Troubleshooting
175
WE14
Exit with vial
information pending
on screen. Exit?
Message appears if you
select Exit without saving
data on Vial Entry, or if you
exit Positive or Negative
Removal display without
scanning a pulled vial or
confirming the sequence
number of a manually entered
vial.
Select Yes to exit without saving the
data.
Select No to return to the display with
data retained on the display.
Then select Save to save the data.
WE16
Only one drawer
may be open at a
time while
performing vial
workflows.
You opened a second drawer. Only one drawer can be open for vial
entry /removal or maintenance
activities.
WE17
Sequence scanned
belongs to vial in
station above.
Consult Manual!
A known vial sequence
number is scanned for a vial
currently in the instrument.
Vial may have been removed
when drawer was offline.
Vials may have been
swapped.
The sequence and station are
shown at the top of the
message box.
For optimal recovery, subculture both
vials (the scanned vial and the one in
the station named in the message).
You may also apply a replacement
barcode to either or both vials and
reenter them with Vial Entry to
continue testing.
WE20
Accession does not
match previously
associated
accession.
Accession can be
disassociated at
Culture Screen.
During Vial Entry or ID
Anonymous, you scan or
enter an accession and vial
sequence number, but the
sequence belongs to a
different accession.
To change the accession, go to Culture
– Vial display and disassociate the vial
from the accession number.
Then enter the correct accession
number.
Select OK to remove the message
box.
WE21
Vial sequence is a
replacement
barcode. Select a
medium type.
Replacement vial barcode
labels have a generic medium
type of 99.
The system performs
optimally when the correct
medium type is known for a
given vial.
Select the medium type by selecting
the Media field and selecting the
correct medium type in the dropdown
box.
Select OK to remove the message
box.
WE24
Remove any vial
from station and
insert plug.
Appears when an empty
station is blocked using
Block/Unblock utility.
Insert a station plug to prevent
inserting a vial in the blocked station.
Select OK to remove the message
box.
Error
No.
Message Possible Cause(s) Corrective Action(s)