2 – Installation
23
2.2 Instrument Specifications
Physical Dimensions Single Instrument Stack
Height 93.9 cm (37.0 in) 198.7 cm (78.25 in)
Width 62.2 cm (24.5 in) 62.2 cm (24.5 in)
Depth 86.9 cm (34.25 in) 86.9 cm (34.25 in)
Clearance (rear, left, right) 1.3 cm, 0 cm, 0 cm 1.3 cm, 0 cm, 0 cm
Clearance (front) 68.6 cm (27.0 in) 68.6 cm (27.0 in)
Weight (empty) 187.5 kg (413.4 lb) 384.8 kg (848.4 lb)
Weight (full) 220.4 kg (485.9 lb) 451 kg (994.2 lb)
Stand weight 63.5 kg (140 lb) N/A
Counterweights (countertop, unanchored) 47.6 kg (105 lb) N/A
Electrical Requirements
Input Voltage 100–240 VAC ± 10%
Peak Current 8 amperes
Input Line Frequency 50/60 Hz
Power 800 W
Heat 450 Btu/hr
Environmental Requirements
Non-Operating Storage
Temperature -17.8 to 65 °C (0–149 °F)
Humidity 10–90% RH, non-condensing
Operating Conditions
Temperature 18.0–30.0 °C (64.4–86 °F)
Humidity
25–80% RH for temperatures
30.0 °C (86 °F)
Maximum dew point 26.1 °C for temperatures >30.0 °C (86 °F)