BD BACTEC™ FX Instrument User’s Manual
176
WE28
Unexpected vial
pulled during batch
removal.
The system lights all negative
stations when batch removal
is enabled and the Remove
Negative vials activity is
initiated.
If a sequenced vial is
removed from a station that is
not illuminated, this message
appears.
The vial sequence number,
accession, station, and status
are shown in the message
box.
Select OK to remove the message
box.
Use the Vial Entry activity to return the
vial if you did not intend to remove it.
WE29
Anonymous vial
pulled unexpectedly
during batch
removal. Readings
discarded.
The system lights all negative
stations when batch removal
is enabled and the Remove
Negative vials activity is
initiated.
If an anonymous vial is
removed from a station that is
not illuminated, this message
appears.
The station and status are
shown in the message box.
Accumulated test readings are
discarded.
Select OK to remove the message
box.
Note the location and status of the vial
that is displayed at the top of the
message box. Continue removing
negative vials.
Vial should be subcultured and
reentered with the Vial Entry activity.
WE30
Positive vial(s)
present.
Positive vial has been
detected; message appears
when instrument detects first
positive vial in a drawer, when
offline drawer goes online
again, or after power is
cycled.
The drawer is shown at the
top of the message box.
Message is displayed for
each drawer where the first
positive detection occurs.
Select OK to remove the message box
and silence the Positive Alarm tone.
Remove positive vials.
WE31
Drawer contains
sequenced vials
that are in unusable
stations. Consult
Manual. Remove
Vials?
A drawer that contains
sequenced vials in unusable
stations was opened.
Refer to Section 7.3.2 for additional
information.
WE34
Drawer selected is
currently offline.
Information shown
may not be
up-to-date.
Appears if a drawer that is
offline is selected in Drawer
View or Block/Unblock
Stations display.
The drawer, a station, or a
row board in a drawer could
be what is offline.
Message is informational. Vial or
station statuses may be different from
what is shown on the display because
the instrument cannot communicate
with the offline drawer.
Error
No.
Message Possible Cause(s) Corrective Action(s)