7 – Troubleshooting
177
WE35
Report size
estimate is greater
than 50 pages.
Print Report?
The instrument has
calculated the approximate
size of the report to be more
than 50 pages.
To accept the message, select Yes.
Select No to cancel the print request.
WE36
Drawer contains
anonymous vials
that are in unusable
stations. Print
Affected Report and
Identify vials
through Id
Anonymous
workflow. Consult
Manual.
A drawer that contains
anonymous vials in unusable
stations was opened.
Use the ID Anonymous activity to
identify any vials in the drawer. When
identifying anonymous vials in this
scenario, be sure to either select Save
after identification to move them to
another drawer, or place the vials in
another station in the drawer that is lit
steady green.
WE53
Drawer contains
one or more vials
that are partially
seated. Fully insert
Vials!
The instrument has
determined that a vial may be
partially seated in the station.
Message is displayed when
the drawer is first opened and
each time it is subsequently
opened, until a measurement
occurs that clears the partial
insertion condition.
Refer to Section 7.3.4 for additional
information.
WE56
One or more
drawers is ajar.
Please close or fully
open any
drawer(s).
The drawer sensor has
detected that the drawer is
not fully closed.
Push drawer fully closed.
WE57
Database under-
write
– displayed
data not current.
Please change data
again and
reattempt save.
While you were attempting to
enter a new vial, identify an
anonymous vial, or change
vial or specimen information,
another process changed
information for that vial or
specimen.
Your current modifications are not
saved.
On the Culture display, recall the vial/
accession and modify the desired
information again.
WE58
Vial can no longer
be found in the
database and
cannot be viewed
or modified.
Consult manual.
This will be displayed if the
user has attempted to modify
a vial that has been deleted
by BD EpiCenter since the
time it was recalled on the
Culture screen.
Message is informational. No activity is
possible since the vial is no longer
resident in the database.
Error
No.
Message Possible Cause(s) Corrective Action(s)