BD BACTEC™ FX Instrument User’s Manual
178
WE59
Vial last known to
be in station shown
above, which is
offline. Is this the
same vial?
Sequence vial is removed
from a degraded mode
instrument or offline row and
inserted via Vial Entry into an
online instrument.
Select Yes to remove the vial from the
old (offline) location.
Place the vial into the online
instrument to continue testing the vial
in its current protocol.
Select No to re-enter the vial sequence
number (in case sequence was
entered incorrectly).
WE60
Vial currently
residing in an offline
instrument and
cannot be modified.
Consult manual.
In a BD EpiCenter
configuration, you tried to
modify a vial in the Culture
display (e.g., disassociate or
associate accession, change
protocol length, select media
type (if replacement vial),
change status) if the vial
resides in an instrument
which is currently offline.
You cannot modify information for a
vial that is in an offline instrument.
WE61
Vial removed from
station shown
above is positive.
Message occurs when a
removed vial is called
Positive. This will likely be the
result of an anonymous vial
being identified and saved
(i.e. not returned to the
instrument). Algorithms are
re-run on the vial with the
newly identified media type,
and if more sensitive than the
general algorithm, could
cause a positive result.
Message is informational. Vial is
already removed from the instrument.
WE62
Restoring
connection to
database. Please
wait
The instrument and
BD
EpiCenter systems are
reconciling their databases.
Message is displayed while the
reconciliation process is in progress.
Message is removed when the
reconciliation process is complete.
WE63
Drawer is
experiencing an
agitation problem
and cannot take
readings.
Four retry attempts to resume
agitation have failed. (See
System Alert 37)
Unrecoverable issue with
agitation system.
Move all vials to another drawer.
Contact BD.
Error
No.
Message Possible Cause(s) Corrective Action(s)