EasyManua.ls Logo

BD BACTEC FX - Page 72

Default Icon
204 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BD BACTEC™ FX Instrument User’s Manual
72
Use the Vial Entry activity to reenter vials. If the vial is still in the database, the Vial Entry display
shows the existing information, including the previous station. A reentered vial should be placed in
its previous station, which illuminates FLASHING GREEN (if the drawer is open and the station is
unoccupied).
Figure 4-7 – Modify Protocol Window
When reentering a NEGATIVE vial that has been out of the instrument for less than 5 hours, the
Time the
vial is out
of the
instrument
Does BD
recommend
subculturing
before vial
re-entry?
Is positivity analysis
restarted?
Is protocol restarted?
20
minutes.
No
If re-entered
into the
same
workgroup
No If same workgroup No
If moved to
a different
workgroup
Yes
If different
workgroup
If BD EpiCenter Yes
If BD Synapsys No
>20 minutes
and <5
hours.
Yes Yes
If same workgroup No
If different
workgroup
If BD EpiCenter Yes
If BD Synapsys™ No
5 hours.
Yes Yes
If same workgroup Yes
If different
workgroup
If BD EpiCenter Yes
If BD Synapsys Yes

Table of Contents

Related product manuals