5 – Reference
91
Figure 5-4 – ID Anonymous Display
Last Location
This field shows the station number from which the anonymous vial was removed.
TIP or TTD
Read-only field that shows the Time in Protocol (for ongoing vials) or Time to Detection (for
positive vials), in days; hours: minutes.
Time in Protocol is calculated as the amount of time between the vial’s Start of Protocol and
the current time (if in the instrument) or removal time (if removed from the instrument).
Time to Detection is calculated as the amount of time between the vial’s Start of Protocol and
when it is declared positive by the instrument.
Drawer X (where X is the current drawer)
Counters for positive vials (+), negative vials (–), blocked/unusable stations (), ongoing vials
(), anonymous vials (?), and available stations (O).
ID Anonymous Buttons:
Modify button
Select to bring up a window enabling you to change the vial's protocol length. This changes
the protocol only for the current vial. To change the default protocol for a medium type, go to
the Configuration – Lab display.