Monthly Maintenance
Replace Alkaline Buffer
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-27
10
5 Separate the electrode (3), retainer (2) and retainer nut (1). (Refer to Figure
10.14.)
Note: It is not required to clean the stir bar, however, the stir bar may be
cleaned if desired while the BUNm electrode is removed from the cup.
Refer to Four Month Maintenance, "Cleaning MC Reagent Lines", section
in this chapter.
6 Remove the quad-ring (4) from the front face of the electrode.
CAUTION
Avoid scratching gold coating on face of electrode as excessive
scratching may adversely affect performance.
Figure 10.14
7 Rinse the quad-ring in deionized water. Dry using lintless tissue.
Step Action, continued
(3 of 5)