Six-Month Maintenance
Replace Calcium and Potassium Electrode Tip
UniCel DxC Systems Instructions For Use A13914 Maintenance
October 2005 Page 10-57
10
ISE Service
The ISE service procedure prevents reagent leakage from the flow cell whenever
electrode removal is required.
Step Action
1 Loosen the two Phillips screws (1) on the ISE module cover and lift off
cover.
Figure 10.34
2 Release the locking pin (1) on the ISE module and lift the module until it
clicks. The locking pin is located on the left side of the ISE module on
DxC 800 Systems as shown below. The locking pin is located on the right
side of the ISE module on DxC 600 Systems.
Figure 10.35
(1 of 2)