OPERON D 850 SURGICAL TABLE
SPECIAL SAFETY INSTRUCTIONS
700000120 Rev. 1
S-11
ELECTROMAGNETIC INTERFERENCE
WARNING
WARNING — When using the table
in the vicinity of high-frequency
surgical equipment, such as cardiac
defibrillators and cardiac
defibrillator monitors, refer to the
equipment manufacturers’
instructions to ensure compatibility
issues are considered.
WARNING
WARNING — When using the table
in the vicinity of neuro-diagnostic
monitoring equipment, interference
may appear on the monitoring
device. To eliminate this, set the
BACKLIGHT TIMER to 1 sec using
the service software
ELECTRICAL SYSTEM
WARNING
WARNING – When servicing the
table, turn power off or remove the
appropriate fuses to prevent
personal injury. If power is
required, use care to avoid contact
with live parts.
WARNING
WARNING – Even a “dead”
battery can contain enough power to
cause a serious shock or explosive
spark. Always disconnect the
battery charger plug at the power
center before disconnecting cables.
WARNING
WARNING – The batteries contain
acid, which can cause serious burns.
While the batteries are sealed, a
dead battery may be damaged and
may leak. Always handle batteries
in a way that will prevent acid from
spilling on your skin or clothes.
CLEANING AND DISINFECTION
WARNING
WARNING — Unplug the power
cord before cleaning the table if
cleaning liquids will be used near the
power cord receptacle. Under no
circumstances should steam or
extremely hot water (over
150°F/66°C) be used to disinfect or
clean the table. Elevated
temperatures and high humidity will
weaken the hydraulic lines and result
in failure of the hydraulic system and
increased leakage currents. Do not
clean this table in automatic cart
washing equipment. Use of high-
pressure water guns could lead to an
electrical malfunction and must be
avoided.
CAUTION
CAUTION — Failure to thoroughly
dry the surface after cleaning may
result in rust.
WARNING
WARNING — Always replace fuses
with fuses of the same voltage and
current rating. Failure to use the
correct fuse could cause a fire and
injury to the patient or hospital
personnel.
WARNING LABELS
Figure S-2 shows the patient safety warning
label attached to the table. The label is located
on the base at the foot end.