EasyManua.ls Logo

Berchtold OPERON D 850 - Special Safety Instructions; General Safety Warnings and Symbols; Patient Positioning and Operational Safety

Berchtold OPERON D 850
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERON D 850 SURGICAL TABLE
SPECIAL SAFETY INSTRUCTIONS
Rev. 1 700000120
S-7
IMPORTANT -
THE SAFETY INSTRUCTIONS THAT FOLLOW APPEAR WITHIN THE MANUAL.
READ THEM CAREFULLY BEFORE OPERATING THE UNIT AND FOLLOW
INSTRUCTIONS.
SPECIAL SAFETY INSTRUCTIONS
SAFETY SYMBOLS
This manual uses special symbols to help alert
you to important safety information:
WARNING
This symbol identifies a WARNING.
A warning indicates that the
procedure in the following text
involves actions that could cause
harm or death to an individual if
certain precautions are not taken.
The warning text always provides
specific information on what you
must do to avoid the risk.
CAUTION
This symbol identifies a CAUTION.
A caution indicates that the
procedures in the following text
involve actions that could cause
damage to the equipment, including
the failure to operate. The caution
text always provides specific
information on what you must do to
avoid the risk.
NOTE
This symbol identifies a NOTE. A
note contains information that can
help you perform the procedure in
the following text more effectively.
USE CONDITIONS
WARNING
WARNING — POSSIBLE
EXPLOSION HAZARD: Do not
operate the table in the presence of
flammable anesthetics. The
OPERON D 850 table is designed
for use in non-hazardous
anesthetizing locations.
POWER SUPPLY
WARNING
WARNING — ELECTRICAL
HAZARD: Repeated breaker
tripping or fuse replacement may
indicate a ground fault or overload
condition. The table will not
operate reliably under such a
condition and may be electrically
unsafe. Do not use the table until
this condition is corrected.
LOAD CAPACITY
In normal orientation, this table has a patient
lifting and articulation capacity of 1000 lb
(453 kg). In reverse orientation, this table has
a patient lifting and articulation capacity of
500 lbs (227 kg). Positioning and constraints
are discussed on page S-9. Do not exceed this
maximum load in service. See Figure 1-1.
WARNING
WARNING — Overloading may
result in tipping of the table or
failure of the table sections to
move. Do not exceed the
maximum load in service.
NOTE
NOTE — The use of some
positioning capabilities or table
accessories will put additional
restrictions on patient weight.
Accessories provided by Berchtold
Corp., when used as intended, are
designed to perform safely with
the patient loads for which this
table was designed. A Berchtold
Corporation accessory that puts
additional restrictions on patient
weight, when used as intended,
will be clearly labeled to indicate
the maximum allowed loads.
The table load capacity is 400 lb (181 kg)
when the beach chair accessory is installed on
the table.
PINCH POINTS
WARNING
WARNING — Cutting or crushing
injury may occur if any part of a
patient or staff member is caught in
a pinch point when the table is
articulated.
CAUTION
CAUTION — Damage to the table
or object may occur if any foreign
object is caught in a table pinch
point when the table is articulated.
Due to the variety of table positions attainable
with the OPERON D 850 table, pinch or
crushing points exist (Ref. Figure S-1). The
OR staff must be trained to ensure that body
parts or surgical apparatus are not caught in
pinch points when the table is operated.

Table of Contents