6.2 Software Functions and System Configuration
60
Display per calibration point
- Rate (cps)
Read in the calibration count rate by pressing <run>.
Stop the calibration measurement by pressing <run> again.
Press <enter> to accept the values.
The calibration count rate can also be entered manually.
- Speed (m/s)
Automatic display of the averaged belt speed.
- Measuring time (s)
Display of measuring time.
- Comparative weight (t)
Enter the comparative weight for this calibration point. This value
can be entered before or after the calibration measurement.
Note: After the calibration measurement has been stopped,
confirm the respective measured value with <enter>; oth-
erwise it will be rejected when you quit the respective cali-
bration point.
See detailed discussion in section 5.2.4 b) Calibration on page 44.