CB (E6.1) 07/2017 Page 102/145
Regularly use the hot air sterilization function (chap. 18.3) following cleaning. The shelves and the
emptied water pan can remain inside the chamber during this operation.
Have replaced the CO
2
sterile filter (gas fine filter) once or twice a year.
What to do in case of contamination?
Throw away / autoclave contaminated cultures.
Carefully inspect cultures that seem to be uncontaminated.
Clean the chamber as described. Wipe the inner chamber and the doors with a disinfectant and allow
drying. Autoclave the shelves. Empty the water pan and autoclave it.
Perform a hot air sterilization.