CB (E6.1) 07/2017 Page 41/145
6.2 Operating modes
In the “control off mode (chap. 6.2.1), the controller is non-functional and displays only the actual
values. There is no heating or refrigeration and no inlet of CO
2
or O
2
(chamber with O
2
control). The
temperature approximates the ambient value.
You can enter the desired setpoint values in “fixed value” mode (chap. 7). The controller then operates
as a fixed-point control, i.e., it reaches and maintains the defined setpoints until the next manual change.
6.2.1 Activating the “control off mode or change to “fixed value” operating mode
To select the “control offor “fixed value” operating mode, go to Menu > Controller mode
Fixed value
08.03.2013 05:05:06
General menu.
Select “Controller Mode”
and press the operating button.
..\ Menu
Controller mode
Event list
Alarms
Setpoints
Safety controller
Import/Export
Hot-air sterilization
Humidity control
Close Home
Fixed value
08.03.2013 05:05:06
Submenu “Controller Mode”.
Select the desired controller mode
“Control off” or “Fixed value”
and press the operating button.
..\ Controller mode
Control off
Fixed value
O2 control
Close Home