CB (E6.1) 07/2017 Page 127/145
Fault description / indication
Condensation on the divided
inner door, gas-proof.
Chamber placed on very cold
floor.
Place the chamber on a BINDER
stand to increase the distance to the
Doors not closed properly.
Reduce humidity (chap. 6.4).
Door heating set too low.
Contact BINDER Service to adjust
the door heating
Too low humidity inside.
Increase humidity (chap. 6.4)
Chamber door not properly
Completely close chamber door.
Permadry™ pan empty.
Fill the outer Permadry™ pan with
water up to the marking on the inner
pan with distilled, sterile water. The
pan must have thorough contact to
the bottom of the inner chamber, see
21.7 Controller
Fault description / indication
keypad possible.
Notification
Key lock activated.
Enter the key lock password (chap.
8).
21.8 Open door
A door contact switch serves to check the state of the outer door. When the door is open, temperature,
CO
2
, and O
2
(chamber with O
2
control) control are turned off.
Outer door open
(longer than the
selected alarm delay
time)
Alarm message
Outer door open or not properly
closed.
Confirm the alarm (chap. 11.4). Close
the door. The triggered zero-voltage
relay alarm contact switches back.