CB (E6.1) 07/2017 Page 109/145
If a hot-air sterilization is already running and you try to start it again, the message „Hot air sterilization
running“ is displayed. Confirm the message with “OK”.
Opening the outer door during sterilization leads to aborting the automatic sterilization cycle.
Interruption of the temperature reaction time.
Ineffective sterilization.
∅ Do NOT open the chamber doors during sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅ Do NOT touch the glass door and inner surfaces during sterilization
18.3.3 Aborting the hot-air sterilization – general procedure
Three events lead to aborting the automatic sterilization cycle prematurely:
• Abortion via the controller menu
• Opening the outer door
• Turning off the chamber at the main power switch, or a power failure
To abort the hot-air sterilization cycle via the controller menu, go to Menu > Hot-air sterilization
or Quick menu > Hot-air sterilization
08.02.2017 05:05:06
Submenu “Hot-air sterilization”.
To abort the hot-air sterilization, select “Stop”
and press the operating button.
..\ Hot-air sterilization
Close Home
08.02.2017 05:05:06
Submenu “Stop hot-air sterilization”.
This is a security question.
To abort the hot-air sterilization, select “Really stop”
and press the operating button.
..\ Stop hot-air sterilization
Close Home
The controller returns to the initial view