CB (E6.1) 07/2017 Page 112/145
18.3.5 Terminating successful sterilization during the cooling-down phase (abortion
after more than 4 hours)
The duration of the entire sterilization is approx. 10 hours. If you want to shorten the sterilization
procedure in order to save time, you can abort it during the cooling-down phase, i.e. no sooner than after
4 hours. At this point, the inner temperature is still approx. 140 °C / 284°F.
The effective sterilization phase is automatically finished after 4 hours The chamber is in the cooling-
down phase, meaning that the necessary duration for the proper sterilization phase has occurred.
This symbol on the controller display indicates a successful hot-air sterilization.
The sterilization cycle is now in the cooling down phase.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
Do NOT touch the glass door and inner surfaces for approx. 4 hours after aborting
sterilization.
Fixed value
08.02.2017 05:05:06
Initial view after finishing the sterilization phase
(sample values).
For further information, press the “Info” button.
Setpoint
Actual value
Temperature [°C]
37.0
60.2
CO2 [Vol.-%]
5.0
---
O2 [Vol.-%]
20.0
---
User
Info
Menu
Sterilization information
08.02.2017 05:05:06
Information menu “Sterilization information”.
Confirm the information with “Ok”.
Sterilization information
STERILIZATION FINISHED
Close Ok Home
As the 6-hour cooling-down phase is still running, the symbol continues to be shown on the controller
display. You must now abort the sterilization cycle.