CB (E6.1) 07/2017 Page 126/145
Fault description / indication
No gas control after hot-air
sterilization.
Sterilization cycle still running.
Abort hot-air sterilization.
Information message not
confirmed.
Confirm the information message
with “Ok”.
Inner chamber temperature
Let the inner chamber temperature
21.5 Sterilization
The sterilization is running.
Sterilization cycle is in the
heating up or holding phase
Confirm the alarm (chap. 11.4). Do
not open the door.
“DO NOT OPEN THE
Sterilization phase (4 hours) of
the sterilization is successfully
completed: Interior is sterilized.
Sterilization phase (4 hours) of
the sterilization cycle
completed.
Confirm the information message
(chap. 11.4). The inner chamber
and parts inside can still be hot.
Do not touch.
You can open the doors.
If the sterilization cycle is still running:
abort hot-air sterilization. Let the
chamber cool down (chap. 18.3) and
take it into operation again (chap. 5).
Notification
“STERILIZATION
FINISHED”.
Abortion of sterilization after
less than 4 hours. Interior is not
sterilized.
Abortion of the hot-air
sterilization via the controller
menu or by turning off the
chamber.
Confirm the information message
(chap. 11.4). The inner chamber
and parts inside can still be hot.
Do not touch.
Interior is not sterilized. Start again
the hot-air sterilization or disinfect the
Notification
“STERILIZATION
ABORTED”.
Outer door opened during hot-
air sterilization.
Confirm the information message
(chap. 11.4). Close the chamber door
and start again the hot-air
21.6 Humidity
Fault description / indication
Condensations inside the
chamber.
Permadry™ pan filled with
water when the chamber is not
Empty Permadry™ pan when the
chamber is not operating.
Doors not closed properly.
Reduce humidity (chap. 6.4)
Condensation on the divided
inner door.
Chamber placed on very cold
floor.
Place the chamber on a BINDER
stand to increase the distance to the
Doors not closed properly.
Reduce humidity (chap. 6.4)
Door heating set too low.
Contact BINDER Service to adjust