CB (E6.1) 07/2017 Page 126/145
Fault description / indication
Possible cause
Required measures
No gas control after hot-air
sterilization.
Sterilization cycle still running.
Abort hot-air sterilization.
Information message not
confirmed.
Confirm the information message
with “Ok”.
Inner chamber temperature
above 80 °C
Let the inner chamber temperature
decrease below 80 °C.
21.5 Sterilization
Indication
Possible cause
Required measures
The sterilization is running.
Sterilization cycle is in the
heating up or holding phase
Confirm the alarm (chap. 11.4). Do
not open the door.
Notification
DO NOT OPEN THE
DOOR !”.
Sterilization phase (4 hours) of
the sterilization is successfully
completed: Interior is sterilized.
Sterilization phase (4 hours) of
the sterilization cycle
completed.
Confirm the information message
(chap. 11.4). The inner chamber
and parts inside can still be hot.
Do not touch.
You can open the doors.
If the sterilization cycle is still running:
abort hot-air sterilization. Let the
chamber cool down (chap. 18.3) and
take it into operation again (chap. 5).
Notification
STERILIZATION
FINISHED”.
Abortion of sterilization after
less than 4 hours. Interior is not
sterilized.
Abortion of the hot-air
sterilization via the controller
menu or by turning off the
chamber.
Confirm the information message
(chap. 11.4). The inner chamber
and parts inside can still be hot.
Do not touch.
Interior is not sterilized. Start again
the hot-air sterilization or disinfect the
interior (chap. 18.2)
Notification
STERILIZATION
ABORTED”.
Outer door opened during hot-
air sterilization.
Confirm the information message
(chap. 11.4). Close the chamber door
and start again the hot-air
sterilization.
21.6 Humidity
Fault description / indication
Possible cause
Required measures
Condensations inside the
chamber.
Permadry™ pan filled with
water when the chamber is not
operating.
Empty Permadry™ pan when the
chamber is not operating.
Door gaskets defective.
Replace door gaskets
Doors not closed properly.
Close door properly.
Humidity set too high.
Reduce humidity (chap. 6.4)
Condensation on the divided
inner door.
Chamber placed on very cold
floor.
Place the chamber on a BINDER
stand to increase the distance to the
floor.
Door gaskets defective.
Replace door gaskets
Doors not closed properly.
Close door properly.
Humidity set too high.
Reduce humidity (chap. 6.4)
Door heating set too low.
Contact BINDER Service to adjust
the door heating