CB (E6.1) 07/2017 Page 110/145
Aborting the sterilization cycle means cancelling it before the entire 10-hour duration is completed.
Whether effective sterilization has occurred depends on the time that has elapsed bevor abortion.
• Aborting sterilization after less than 4 hours: Prevents effective sterilization.
This symbol on the controller display indicates that the hot-air sterilization was aborted.
The interior may still be hot. Do not touch the glass door and inner surfaces!
• Aborting / terminating sterilization after more than 4 hours: The chamber is in the cooling-down phase,
meaning that the necessary duration for the proper sterilization phase has occurred.
This symbol on the controller display indicates a successful hot-air sterilization.
The sterilization cycle is now in the cooling down phase.
18.3.4 Aborting hot-air sterilization after less than 4 hours
08.02.2017 05:05:06
Initial view during the hot-air sterilization (sample
values).
For further information, press the “Info” button.
Temperature [°C]
Section 01 End: 08.03.2013 15.48:19
CO2 [Vol.-%]
O2 [Vol.-%]
User
Info
Menu
To abort the sterilization cycle proceed as described in chap. 18.3.3.
This symbol on the controller display indicates that the hot-air sterilization was aborted. The
interior may still be hot. Do not touch the glass door and inner surfaces!
When aborting the sterilization prematurely, it may be that the cells/pathogens inside the chamber have
not all been killed. If necessary, you should repeat the sterilization.
CAUTION
Interruption of temperature reaction time.
Ineffective sterilization.
If necessary, repeat the sterilization.
CAUTION
Glass door and inner chamber become hot during sterilization.
Danger of burning.
∅ Do NOT touch the glass door and inner surfaces for approx. 7 hours after aborting
sterilization.