CB (E6.1) 07/2017 Page 45/145
6.4 Humidity control of the Permadry™ system
The Permadry™ system with its 2-pan water system guarantees a maximum humidity of up to 95 % r.H.
in the inner chamber, which remains condensation-free. This performance assumes an average ambient
temperature of 22 ±3 °C / 71.6 ±5.4 °F and a working temperature in the inner chamber of 37 °C /
98.6 °F.
You can increase or decrease humidity slightly if required.
To configure the humidity control of the Permadry™ system, go to Menu > Humidity control
08.03.2013 05:05:06
Submenu “Humidity control”.
The current setting is displayed.
To change the setting, press the operating button.
Factory default setting: “Medium”.
..\ Humidity control Low/Medium/High
Close Home
08.03.2013 05:05:06
Submenu “Humidity control”.
Select the desired setting
and press the operating button
Close Home
If the ambient temperature deviates by more than +/- 5 °C from the values recommended by
the manufacturer, the conditions for maximum air humidity with condensation-free inner
chamber are no longer guaranteed. Contact BINDER Service for assistance.