CB (E6.1) 07/2017 Page 72/145
Icon
Alarm message
Signification
Door open
Door is open for more than the set alarm delay time (chap. 11.5.5, factory
setting: 1 minute). Close the door.
Low pressure CO2 Low CO
2
outlet pressure (< 0.3 bar)
Low pressure CO2 Low O
2
outlet pressure (< 0.3 bar) (chamber with O
2
control)
Low pressure N2 Low N
2
outlet pressure (< 0.3 bar) (chamber with O
2
control)
CO2 sensor
defective
CO
2
sensor defective. Contact BINDER Service.
O2 sensor
defective
O
2
sensor defective. Contact BINDER Service.
Inform service -
adjust CO2 sensor
CO
2
sensor adjustment required. Contact BINDER Service.
You can activate / deactivate the buzzer in the “Alarms” submenu (chap. 11.5.3).
With an activated buzzer there is an audible alert with an alarm. You can reset it in the “Alarms”
submenu for alarm acknowledgement pressing the “Reset” button (chap. 11.4). The alarm symbol will
only disappear when the cause of the alarm has been remedied.
The zero-voltage relay alarm contact (chap. 11.6) is activated in case of the alarm messages “door
open” and deviations of temperature and gas concentrations) as well as in case of a power failure and
turning off the chamber at its main power switch.
For appropriate actions in the event of an alarm, please refer to chap. 21Troubleshooting”.