Proton Spectrum
58 / 86 H171804E_14_001
Lock the solvent
On the Workflow button bar, click Lock.
In the Solvents table window, select the solvent, e.g. DMSO. Click OK.
Tune and match the probe
On the Workflow button bar, click Tune.
This performs an atma (automatic tuning and matching) and requires a probe equipped
with an automatic tuning module. For more options, click the drop-down arrow on the Tune
button.
Spin the sample (optional)
On the Spin button, click the drop-down arrow to see more options.
In the list, select Turn sample rotation on (ro on).
Rotation may be turned OFF for probes such as BBI, TXI, TBI and for small sample probes.