Safety of CryoProbe equipment
CryoProbe Installation BRUKER 11 (93)
Magnetic stray field
When working within the 0.5 mT stray field of the magnet, all magnetic parts and
tools must be avoided or handled with great care.
Safety of CryoProbe equipment 1.5
0
CAUTION:
- Do not bend the CryoProbe.
Do not hold the CryoProbe at its upper tube, carry it only at its body.
- Do not open the CryoProbe.
There are no user-serviceable parts inside. A CryoProbe cannot be sealed
or reassembled without special equipment. Even undoing some screws can
destroy factory settings and will in general render the CryoProbe unusable.
- Never force a CryoCoupler into position.
- Do not obstruct the operation of the safety-valves on the top and front faces of
the CryoProbe body.
- Do not move a cryogenically cold device.
- Do not try to fix a leak on a cold part because cracking of frozen o-rings, valves
etc. may occur.
- Excessive RF power can destroy the CryoProbe or the HPPR CRP. Obey the
limitations given on the specific ‘L
IMITATIONS - WARNINGS’ sheet. See also
User Manual.
First aid 1.6
0
If cold helium gas comes in contact with eyes or skin, immediately flood the affect-
ed area with cold or tepid water.