70 (93) BRUKER CryoProbe Installation
Relocation of a CryoPlatform
Pressure release 6.3.1
To permanently release the gas pressure from the CryoCooling Unit, shut down
the whole system but keep the Flexlines and the He Compressor connected.
Slowly unlock the pipe connectors marked in Figure 6.1.
and let the gas escape.
Leave the nuts unlocked but keep the connectors on their threads.
Toxic materials 6.3.2
Lead insulation in CryoCooling Unit
The panels of the CryoCooling Unit cabinet contain a lead-enforced polymer foam
for maximum sound insulation. Remove the lead/foam insulation. Recycle or dis-
pose it according to the local regulations for lead-containing materials.
Figure 6.1. Releasing the gas from the CryoCooling Unit