VÅNTEC-1 Detector User Manual Hardware Installation for the D8 ADVANCE
M88-E01072 4 - 5
4.4.2 Install the Controllers
Location of the FDC and PSD controllers is criti-
cal for proper operation. Whenever possible,
place the VÅNTEC-1 controller above the D8
controller electronics (see Figure 4.2 for proper
location in the D8). Allow at least 1" of open
area around the top, bottom, and back of the
units for adequate airflow. Remove the D8 front
and side access panels and blank panels for the
installation of the controllers.
Figure 4.2 - Location of FDC and PSD controller within the
D8 ADVANCE