VÅNTEC-1 Detector User Manual Hardware Installation for the D8 ADVANCE
M88-E01072 4 - 21
4.8 Optic Components
4.8.1 Mount the Detector Head to the
2-Theta Axis
Mount the detector head in place of the scintilla-
tion detector or the previously installed detector.
The detector will be shipped mounted and pre-
aligned to the two degrees of freedom (2 DOF).
If this is not the case, follow the steps indicated
below. Move the detector to different measuring
circles to suit the installation.
1. Move the 2-theta drive to a position where
the detector can be placed on the track
(approximately 35° 2-theta).
2. Check that the setscrew is in the location
shown in Figure 4.21. This ensures that the
detector is at the proper measuring circle.
Figure 4.20 - New detector mount
Figure 4.21 - Detector mounting plate
setscrew
back of detector
front of detector
Detector
mounting
screws