Adjustments and Maintenance
2-20
Step Action Displayed Message Comment
16
Press ESC. Use the
buttons to scroll to PAPER
SENSOR LEVEL (or if
using reflective media,
REFL PAPER LEVEL) and
then press ENTER.
Use the buttons to set the
Paper value determined in
Step 15 and then press
ENTER.
PAPER SENSOR LEVEL
(0 - 255)
173
This is the Paper value.
17
Press ESC. Scroll to GAP
SENSOR LEVEL (or, if
using reflective media,
MARK SENSOR LEVEL)
and then press ENTER.
Set the Gap (or Mark)
value determined in Step
15 and then press ENTER.
GAP SENSOR LEVEL
(0 - 255)
042
This is the Gap (or Mark)
value.
18
Press ESC. Scroll to
EMPTY SENSOR LEVEL
and then press ENTER.
Set the Empty value
determined in Step 15 and
then press ENTER.
EMPTY SENSOR LEVEL
(0 - 255)
009
This is the Empty value.
19
Press EXIT and, when
prompted, press YES at
the SAVE CHANGES
prompt.
READY
The procedure is
finished, the process
almost complete.
20
Load media and position
the Media Sensor. Press
and hold FEED until at
least one label has been
output.
READY
The printer is ready for
use (otherwise, see note
below).
Note: If the Advanced Entry Calibration fails, enter MEDIA SETTINGS CALIBRATION
ADVANCED ENTRY SENSOR GAIN and lower the selected setting number by
one (desensitizing the sensor). Enter the new setting; save the changes; exit
the menu; and, then test your media. Repeat this process until a usable sensor
setting is obtained.