E105, 313327, REV A Protocol 3
TM
User Guide
©2021 ITW EAE. All rights reserved. Despatch is a registered trademark in the U.S. and other
countries. No part of the contents of this manual may be reproduced, copied or transmitted in any form
or by any means without the written permission from ITW EAE, unless for purchaser's personal use.
2.6. Data Recorder
Table 6. Data Recorder.
1Mb non‐volatile flash memory. Data retained when power is
turned off
1; 2; 5; 10; 15; 30 seconds or 1; 2; 5; 10; 15; 30 minutes
Dependent on sample rate and number of values recorded. Two
values can be recorded for up to 7 days at 10s intervals.
More values or faster sample rates reduce the maximum
duration.
VARTA CR 1616 3V Lithium. Clock runs for >1 year without power
(Part # 274030)
Real Time Clock error <1 second per day