116Appendix G: Technical Information
Table 1-A� G6 Accuracy to YSI (n=324)
Patient
Population
Number
of
subjects
Total
number
of paired
CGM-YSI
Percent
within
20/20%
YSI
% (95% LB)
Day 1
Percent
within
20/20%
YSI
MARD
(%)
Overall 324 25,101 91.7 (90.6) 87. 8 9.8
Adults
(18+ YO)
159 19,329 91.6 (90.3) 87.1 9.9
Pediatrics
(6-17 YO)
165 5,772 92.0 (89.8) 90.2 9.6
Pediatrics
(2-5 YO)*
8 82 92.7 (86.6) NA 9.9
* No YSI measurements were taken for this age group; results
presented are from in-clinic CGM-SMBG matched paired
measurements.
1