16Chapter 2: Safety Statements
Clean insertion site with alcohol wipes to prevent infections. Don’t
insert the sensor until skin is dry. If insertion site is not clean and
completely dry, the patient runs the risk of infection or the transmitter
holder not sticking well.
Make sure patient doesn’t have insect repellent, sunscreen, perfume,
or lotion on their skin.
Where to Insert: Things to Check
Don’t remove sensor or transmitter from packaging until you are ready
to use. Keep the safety guard on until you put the G6 Pro applicator
against the patient’s skin. If you remove the safety guard first, you may
hurt the patient by accidentally pushing the button that inserts the
sensor before you mean to.
Change the insertion site with each sensor. Using the same site too
often on the same patient might not allow the skin to heal, causing
scarring or skin irritation.
Sensor placement is important. Choose a site:
At least 3 inches from insulin pump infusion set or injection site
Away from waistband, scarring, tattoos, irritation, and bones
Unlikely to be bumped, pushed, or laid on while sleeping
Follow G6 Pro instructions. If not, your patient could have a severe low
or high glucose event.
Use Correct Transmitter, and Sensor
G6 Pro components are not compatible with any previous
Dexcom products. Do not mix transmitters and sensors from
different generations.

Table of Contents