128Appendix G: Technical Information
Table 4-B� Trend Accuracy Rate of Change (Pediatrics*; n=165)
CGM Rate
Range
(mg/dL/
min)
YSI glucose range (mg/dL)
CGM-YSI
Pairs
(n)
<-2 [-2,-1) [-1,-0) [0,1] (1,2] >2
<-2 47.9% 37.0% 12.8% 1.9% 0.0% 0.5% 211
[-2,-1) 6.6% 55.5% 33.8% 3.4% 0.6% 0.1% 686
[-1,0) 0.5% 8.9% 73.7% 15.8% 1.0% 0.0% 2,048
[0,1] 0.0% 0.8% 25.5% 62.9% 10.0% 0.8% 1,666
(1,2] 0.0% 0.4% 4.4% 35.9% 48.0% 11.4% 546
>2 0.0% 0.5% 1.7% 7.1% 23.6% 67.1 % 423
* Includes pediatric subjects 6-17 years of age; no YSI measurements
were taken for pediatric subjects 2-5 years of age.
Hypoglycemia and Hyperglycemia Alerts
Low and High Glucose Alerts
The ability of the G6 to detect high and low glucose levels is assessed
by comparing G6 results to YSI results at low and high blood glucose
levels and determining if the alert may have sounded. The G6 and YSI
values were compared by pairing the G6 Pro reading and the YSI value
within before or after 15 minutes of each other. We suggest that you
ask your doctor what alert settings would be best for you.