G6 Pro User Guide
148Appendix G: Technical Information
Electromagnetic Immunity and Emissions Declaration and
Guidance
The transmitter is intended for use in the electromagnetic environment
specified in the next table. The customer or the user of the transmitter
should ensure that it is used in such an environment.
Immunity Test Transmitter Compliance Level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field
(50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
N/A
Surge
IEC 61000-4-5
N/A
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
N/A
Conducted Fields
Disturbance
IEC 61000-4-6
N/A