EasyManua.ls Logo

Dexcom G6 Pro - Sensor Stability

Dexcom G6 Pro
172 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
G6 Pro User Guide
135Appendix G: Technical Information
Sensor Stability
Sensors can be worn for up to 10 days. Performance was estimated by
calculating the percentage of G6 Pro readings within 15 mg/dL or 15%
(15/15%), 20 mg/dL or 20% (20/20%), and 40 mg/dL or 40% (40/40%),
of the YSI values at the beginning (Day 1, 2), middle (Day 4, 5) and
end (Day 7, 10) of the G6 lifecycle. For blood glucose values less than
or equal to 70 mg/dL, the absolute difference in mg/dL between the
two glucose results was calculated. For values greater than 70 mg/dL,
the absolute difference (%) relative to the YSI values was calculated.
In addition, the mean absolute relative difference (MARD) shows the
average amount the sensor readings differ from the YSI glucose. The
MARD values included in Table 6-A and 6-B show consistent accuracy
and sensor stability over the 10-day life of the sensor.
Table 6-A� Sensor Stability Relative to YSI (Accuracy over Time
1
)
(Adults; n=159)
Wear
Period
Number
of
paired
CGM-
YSI
MARD
(%)
Percent
within
15/15% YSI
(%)
Percent
within
20/20% YSI
(%)
Percent
within
40/40% YSI
(%)
Beginning 6,696 10.9 76.5 88.0 99.6
Middle 6,464 9.2 84.3 94.6 99.8
End 6,169 9.6 82.3 92.4 99.8
1
CGM readings are within 40-400 mg/dL, inclusive.

Table of Contents

Related product manuals