G6 Pro User Guide
141Appendix G: Technical Information
Number of Readings Provided
The G6 is capable of providing a reading every 5 minutes, or up to
288 readings per day. For a variety of reasons, the G6 may not display
a glucose reading and readings are “skipped.” The percentage of
readings you can expect to receive from the G6 over the sensor life is
98.6%. More than 97% of the sensors captured readings at least 90%
of the time. For the G6 with auto-applicator, approximately 99% of the
sensors displayed reading every 5 minutes at least 90% of the time.
Table 8 below describes the reading captured rate by each wear day
over the sensor life.
Table 8� Reading Capture Rate by Wear Day (n=374)
Wear Day Number of Sensors Capture Rate (%)
1 374 97.6
2 368 98.6
3 364 98.7
4 354 98.6
5 348 98.5
6 338 98.5
7 329 98.2
8 314 97.8
9 303 97.0
10 290 96.4