125Appendix G: Technical Information
Table 3-C� Concurrence of G6 CGM Readings and YSI Values by YSI
Glucose Range (Adults, n=159)
CGM
Glucose
Range
1
(mg/dL)
YSI glucose range (mg/dL)
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
> 400
<40 51.9% 5.0% 1.1% 0.1% 0.1% . . . . . .
40- 60 40.7% 52.7% 11.7% 0.7% 0.1% 0.0% . . . . .
61- 80 7.4% 41.0% 63.7% 11.0% 0.2% 0.1% . . . . .
81- 120 . 1.3% 23.4% 75.8% 19.7% 1.0% 0.0% . . . .
121- 160 . . 0.0% 12.2% 66.9% 24.8% 1.4% 0.0% 0.1% . .
161- 200 . . . 0.1% 13.0% 59.9% 25.3% 1.7% 0.4% 0.2% .
201- 250 . . . . 0.2% 14.1% 61.9% 30.6% 5.1% 0.2% .
251- 300 . . . . . 0.1% 11.3% 56.2% 35.9% 9.6% .
301- 350 . . . . . . 0.1% 11.3% 48.0% 38.0% 26.1%
351- 400 . . . . . . . 0.1% 10.5% 46.0% 47.8%
>400 . . . . . . . . 0.1% 5.9% 26.1%
Total 27 1,180 2,191 3,503 2,910 2,457 2,755 2,383 1,601 437 23
1
CGM readings are within 40 to 400 mg/dL, inclusive.