Instructions for use Fabius plus SW 3.n 101
Operation
Ventilation mode Pressure Control
(optional)
Changing to Pressure Control mode
In the following examples and illustrations, the
change from Volume Control mode to Pressure
Control mode is described:
1 Press the Pressure Control (A) key.
2 Adjust ventilation settings (B).
3 Confirm the new mode.
Due to the influence of compliance and resistance,
the set value for Freq Min in Pressure Control
mode might not exactly be applied.
In the following table, all parameters (B) in
Pressure Control mode, together with the
corresponding setting ranges and factory settings,
are listed.
2547225473
A
B
Volume Control
Pressure Control
Parameter Setting range Factory set-
ting
Inspiratory
pressure
PINSP
[cmH
2O]
([hPa])
5 to 65,
min. PEEP +5
15
Respiratory
rate Freq [bpm]
([1/min])
4 to 60 12
Inspiratory
time:expiratory
time TI:TE
4:1 to 1:4 1:2
Inspiratory flow
Insp Flow
[L/min]
10 to 75 30
PEEP [cmH
2O]
([hPa])
0 to 20 0