Instructions for use Fabius plus SW 3.n 79
Assembly and preparation
Connecting the sensors and measurement lines
Connecting the O2 sensor
1 Insert the O2 sensor (A) into the opening of the
inspiratory valve on the compact breathing
system.
2 Connect the plug (B) of the O
2 sensor cable
with the connection marked with on the
rear of the device.
Connecting the pressure sensor
1 Connect the pressure measurement hose (C)
to the corresponding connection on the bottom
of the compact breathing system.
Make sure that the pressure measurement
hose is not pinched.
2 Connect the other end of the pressure
measurement hose with the bacterial filter (B).
3 Plug in the bacterial filter into the connection
(C) marked with on the rear of the device.