Instructions for use Fabius plus SW 3.n 31
Overview
Supplemental O2 delivery (optional)
The supplemental O2 delivery supplies pure
oxygen with an exact metered flow, e.g., for O
2
insufflation using a face mask during a regional
anesthesia. The supplemental O
2 delivery is not
only possible in standby mode and during
operation, but also if Fabius is switched off.
The supplemental O
2 delivery can supply
additional inspiratory oxygen for the patient for the
following types of anesthesia:
Spinal anesthesia
Epidural anesthesia
Other regional anesthesia
To increase the O
2 concentration in the breathing
gas, the supplemental O
2 delivery can be used in
combination with a breathing bag
*
.
Functional check of the supplemental
O
2 delivery
Turn the flow control valve (A)
counterclockwise.
Check that the float can move freely in the flow
tube.
After the O
2 insufflation is ended, the flow control
valve of the supplementary O
2 delivery must be
completely closed:
Turn the flow control valve (A) clockwise to the
final position stop.
WARNING
Risk due to overpressure
When the patient's connection to the
supplemental O
2 delivery is made using a
breathing circuit without relief valve,
increased pressure may be applied to the
patient.
When connecting the patient, only use a
breathing circuit with relief valve or do not
connect pressure-tight.
WARNING
Risk of fire
The oxygen can ignite when cauterizing near
an oxygen source.
Make sure that all connections (e.g., Y-
piece, breathing hoses) do not leak.
Before cauterizing, close the flow control
valve.
Remove mask.
Wait a few moments.
25317
O
2
Fabius plus
A
* ASTM F1850-22(2005) §76