Instructions for use Fabius plus SW 3.n 105
Operation
Ventilation mode SIMV/PS (optional)
The ventilation mode SIMV (synchronized
intermittent mandatory ventilation) is a mixture of
ventilation and spontaneous breathing. In
ventilation SIMV mode, the patient can breathe
spontaneously. The ventilation is done
synchronously to the inspiratory effort of the
patient.
The mandatory breaths are defined based on the
following parameters:
VT
Freq
TINSP
TIP:TI
PEEP
To support the inspiratory effort of the patient in
ventilation SIMV mode, pressure support PPS
can be switched on. The setting of PPS to
another value than OFF activates the mode
Pressure Support, see chapter ''Ventilation mode
Pressure Support (optional)'' on page 103.
Changing to SIMV mode
In the following examples and illustrations, the
change from Volume Control mode to SIMV/PS
mode is described:
1 Press the SIMV/PS (A) key.
2 Adjust ventilation settings (B).
To set the following additional parameters,
press the MORE (C) key:
Trigger
Insp Flow
TINSP
TIP:TI
3 Confirm the new mode.
2547725478
A
B
Volume Control
SIMV/PS

Table of Contents

Other manuals for Dräger Fabius plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus and is the answer not in the manual?

Dräger Fabius plus Specifications

General IconGeneral
BrandDräger
ModelFabius plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals