Instructions for use Fabius plus SW 3.n 207
Technical data
S-ORC (Sensitive Oxygen Ratio Controller)
Device outlets
At a flow of approx. 200 mL Set the N2O concentration in the fresh gas
between 0 and 75 %.
In case of insufficient O
2 S-ORC limits the N2O concentration in the fresh
gas so that the O
2 concentration does not drop
below 23 Vol%.
N
2O flow control valve is open and at the same
time the O
2 flow control valve is closed or set at
less than 0.2 L/min.
S-ORC prevents N
2O flow.
In case of N2O failure O2 can continue to be supplied. No alarm.
Serial interfaces COM 1
Only connect to devices that meet the require-
ments of IEC 60950-1 for ungrounded SELV
circuits and the requirements of IEC 60601-1
(as of the 2nd edition) for exposed secondary
circuits with maximum 24 Vdc nominal voltage.
Protocol Vitalink, MEDIBUS
Connector 9-pole Sub-D, galvanically isolated with 1.5 kV
against internal electronics, 0.5 kV against
housing
Baud rate 1200, 2400, 4800, 9600, 19200, 38400 baud
Data bits 7 or 8
Parity Uneven, even, none
Start bit 1
Stop bit 1 or 2