Operation
O2 flush
The O2 flush is used for flushing and quickly filling
the breathing system and breathing bag with O
2
while bypassing the vaporizer.
During this an unmetered flow of at least 35 L/min
is given to the breathing system and the breathing
bag.
Press the (A) key. O
2 flows for as long as
the key is held down.
Use of the O
2 flush can increase airway pressure
very quickly and abruptly change the gas
concentration.
20960
NOTE
In ManSpont mode, the pressure can increase
quickly and thus trigger the APL valve.
+