Configuration
150 Instructions for use Fabius plus SW 3.n
Switching the desflurane compensation
on and off
The desflurane compensation optimizes the
volume measurement when desflurane is used.
1 Press the DesComp OFF softkey (A).
The key label changes from DesComp OFF to
DesComp ON.
The desflurane compensation is activated.
The message Des on is displayed in the status bar
(B).
Automatic desflurane compensation
Prerequisite: External gas analyzer is connected to
Fabius via the RS232 interface.
25505
CAUTION
Risk of inaccurate measured values
If the activation of the desflurane compensation
is forgotten during the use of desflurane or if the
desflurane compensation is activated even
though desflurane is not being used, the
accuracy of the volume measurement can be
influenced.
Only switch on desflurane compensation if
desflurane is used.
CAUTION
Risk of inaccurate measured values
If an anesthetic gas monitor is used, the
automatic desflurane compensation is activated.
Faulty anesthetic gas monitors can influence the
accuracy of the measured volume.
Make sure that the anesthetic gas monitor
functions correctly.
B
A
Volume Control
CAUTION
Risk of inaccurate measured values
Desflurane influences the measurement
accuracy of the flow sensor.
If desflurane is used, activate the desflurane
compensation.
25507
NOTE
If data for anesthetic gas concentration is
available via the communication with an external
gas analyzer, Fabius compensates desflurane
automatically. In this case, the transmitted data
cancel the function of the softkey for desflurane
compensation.
A
B
Volume Control

Table of Contents

Other manuals for Dräger Fabius plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Fabius plus and is the answer not in the manual?

Dräger Fabius plus Specifications

General IconGeneral
BrandDräger
ModelFabius plus
CategoryMedical Equipment
LanguageEnglish

Related product manuals