Troubleshooting
Systematic localization of leakages
To find causes of leakages, isolate individual
components from the leakage test.
1 Perform the leakage test, see chapter
''Leakage test'' on page 135.
2 Contact service personnel if the leakages
cannot be localized.
Component Measure
Sample line Remove the sample line.
Occlude the Luer Lock con-
nector on the Y-piece.
Breathing hoses Disconnect the breathing
hoses. Connect the inspiratory
port and expiratory port with a
hose that is known to be with-
out leakages. Connect the
breathing bag directly to the
breathing system.
Vaporizers Remove the vaporizers.