Instructions for use Fabius plus SW 3.n 131
Configuration
Configuration in standby mode
The following configuration functions are available
in Standby mode:
Calibrations
System tests
Management of default settings
1 Press the (A) key.
The pressure waveform window is replaced by a
confirmation message (B) and instructions to turn
off the flow.
The LED of the key (A) starts flashing. It
flashes until the Standby mode is confirmed.
2 Confirm the new mode.
The ventilator changes to Standby mode. The
previous screen is replaced by the start screen
Standby and the standby LED is continually lit.
The following softkeys (C) are displayed on the
start screen:
2541125491
NOTE
If the confirmation is not done within 15 seconds,
the ventilator remains in the previous mode and
the pressure waveform window is restored.
A
B
ManSpont
25463
Run System Test
Calibrate Flow Sensor
Calibrate O
2 Sensor
Leak / Compl Test
Access Alarm Log
Restore Site Defaults
C