Operating concept
42 Instructions for use Fabius plus SW 3.n
Control panel
The control panel contains the following main
elements:
A Screen
B Key to select ventilation modes
– Volume Control
– Pressure Control
– Pressure Support
*
– SIMV/PS *
– ManSpont
C Key for more functions
25357
*optional
Key Function
Opens the window with alarm limits.
In the mode Standby:
Opens a menu for configuring the system
settings and default settings, see chapter
''Configuration in standby mode'' on page
131
In a ventilation mode:
Opens a menu for displaying and chang-
ing the monitoring settings, see chapter
''Configuration during operation'' on page
147
Changes from displayed screen to the
main screen.
The alarm tone of all active alarms is sup-
pressed for 2 minutes.
Switches to Standby mode. Monitoring is
switched off and the ventilator stops.
Key Function