Instructions for use Fabius plus SW 3.n 59
Assembly and preparation
Mounting the anesthetic vaporizers
Depending on the configuration, Fabius can be
operated with vaporizers that have a plug-in
adapter for the following connectors:
–Selectatec
Interlock S
Dräger Auto Exclusion
The anesthetic vaporizers must be mounted in
accordance with the respective instructions for
use.
The vaporizers used must conform to the
applicable safety standard.
If an independent gas measurement system is
used, it must conform to the applicable safety
standard.
WARNING
Risk due to incorrect anesthetic agent
delivery
If the vaporizer is filled with the wrong
anesthetic agent or if it is not filled
sufficiently, incorrect anesthetic gas
concentrations or concentrations that are too
low can occur as a result.
Strictly observe the instructions for use of
the vaporizer.
Compare the color coding on the
vaporizer with the anesthetic agent bottle.
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause
leakage. This can cause the fresh-gas flow to
be too low or contaminate the ambient air.
Patient and user can be endangered.
Make sure that the vaporizers are
mounted levelly.
When using D-Vapor vaporizers, make
sure that the power cable is not pinched.
After mounting the vaporizers, perform a
leakage test.